Remote control commands
R&S
®
LCX Series
187User Manual 1179.2260.02 ─ 02
Example: Managing files
The example represents the remote control commands to manage files in directories.
**************************************************
// Request information on files
// and their contents:
// list of files in a directory,number of measurement
// points in a file, the saved readings.
// Remove a file
DATA:LIST?
DATA:POINts?
DATA:DATA? "/USB1A/LCX100/logging/log-20210221T123456.789.csv"
**************************************************
// Remove a file
DATA:DELete "/USB1A/LCX100/logging/log-20210221T123456.789.csv"
DATA:DATA?................................................................................................................. 187
DATA:DELete................................................................................................................ 187
DATA:LIST?...................................................................................................................187
DATA:POINts?...............................................................................................................188
DATA:DATA? <File Path>
Queries the contents of a file, e.g. the data of a logging file.
Query parameters:
<File Path>
Example: See Chapter 11.7, "DATA subsystem", on page 186.
Usage: Query only
DATA:DELete <File Path>
Removes a file from the specified directory.
Setting parameters:
<File Path>
Example: See Chapter 11.7, "DATA subsystem", on page 186.
Usage: Setting only
Manual operation: See "Delete" on page 114
DATA:LIST?
Queries all files in a specified directory.
Example:
See Chapter 11.7, "DATA subsystem", on page 186.
DATA subsystem