Remote Commands Reference
R&S
®
RTC1000
249User Manual 1335.7352.02 ─ 02
GENerator:VOLTage:OFFSet <Offset>
Sets the DC offset of the selected generator function.
Parameters:
<Offset> Range: -3.00000E+00 to 3.00000E+00
*RST: 0.00E+00
Default unit: V
GENerator:FREQuency <Frequency>
Defines the frequency value of the selected generator function.
Parameters:
<Frequency> Numeric value depending on the selected generator function.
Range: 1.000E-01 to 5.000000000E+04 (sine / square)
1.000E-01 to 1.000000000E+04 (triangle / ramp /
pulse)
*RST: 1.00000000E+03
Default unit: Hz
GENerator:FUNCtion:RAMP:POLarity <Polarity>
Sets the polarity of the generator function ramp.
Parameters:
<Polarity> POSitive | NEGative
*RST: NEG
GENerator:FUNCtion:PULSe:DCYCle <DutyCycle>
Defines the duty cycle value of the generator function pulse.
Parameters:
<DutyCycle> Range: 1.000E+01 to 9.000E+01
*RST: 2.500E+01
Default unit: %
GENerator:OUTPut[:ENABle] <OutputEnable>
Activates or deactivates the function generator output.
Parameters:
<OutputEnable> ON | OFF
*RST: OFF
Function Generator