Remote Commands Reference
R&S
®
RTC1000
172User Manual 1335.7352.02 ─ 02
RUNContinous
Same as RUN.
Usage: Event
Asynchronous command
SINGle
Performs a single acquisition and returns to stop mode.
Usage: Event
Asynchronous command
RUNSingle
Same as SINGle.
Usage:
Event
Asynchronous command
STOP
Stops the running acquistion.
Usage:
Event
Asynchronous command
13.3.2 Time Base
TIMebase:SCALe...........................................................................................................172
TIMebase:RATime?....................................................................................................... 173
TIMebase:ACQTime.......................................................................................................173
TIMebase:RANGe..........................................................................................................173
TIMebase:DIVisions?..................................................................................................... 173
TIMebase:POSition........................................................................................................ 173
TIMebase:REFerence.....................................................................................................174
TIMebase:SCALe <TimeScale>
Sets the horizontal scale for all channel and math waveforms.
Parameters:
<TimeScale> Range: 1e-9 to 50
*RST: 100e-6
Default unit: s/div
Acquisition and Setup

Table of Contents

Other manuals for R&S RTC1000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S RTC1000 and is the answer not in the manual?

R&S RTC1000 Specifications

General IconGeneral
BrandR&S
ModelRTC1000
CategoryTest Equipment
LanguageEnglish

Related product manuals