TIMebase:RATime?
Queries the real acquisition time used in the hardware. If FFT analysis is performed,
the value can differ from the adjusted acquisition time (TIMebase:ACQTime).
Return values:
<HWAcqTime> Range: Depends on various settings
Default unit: s
Usage: Query only
TIMebase:ACQTime
<AcquisitionTime>
Defines the time of one acquisition, that is the time across the 12 divisions of the dia-
gram: Timebase Scale*12.
Parameters:
<AcquisitionTime> Range: 12 ns to 600 s
Default unit: s
TIMebase:RANGe <AcquisitionTime>
Defines the time of one acquisition, that is the time across the 12 divisions of the dia-
gram: Timebase Scale*12.
Parameters:
<AcquisitionTime> Range: 12 ns to 600 s
Default unit: s
TIMebase:DIVisions?
Queries the number of horizontal divisions on the screen.
Return values:
<HorizDivCount> Range: 12
Usage: Query only
TIMebase:POSition <Offset>
Defines the trigger position (trigger offset) - the time interval between trigger point and
reference point to analyze the signal some time before or after the trigger event. If the
trigger position is zero, the trigger point matches the reference point.
See also: TIMebase:REFerence
Parameters:
<Offset> Range: -500 to 500
*RST: 0
Default unit: s
Acquisition and Setup

Table of Contents

Other manuals for R&S RTC1000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S RTC1000 and is the answer not in the manual?

R&S RTC1000 Specifications

General IconGeneral
BrandR&S
ModelRTC1000
CategoryTest Equipment
LanguageEnglish

Related product manuals