EasyManua.ls Logo

R&S RTC1000 - Page 330

R&S RTC1000
371 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands Reference
R&S
®
RTC1000
330User Manual 1335.7352.02 ─ 02
Parameters:
<Setup> 488.2 block data
Usage: SCPI confirmed
SYSTem:ERRor[:NEXT]?
Queries the error/event queue for the oldest item and removes it from the queue. The
response consists of an error number and a short description of the error. Positive
error numbers are instrument-dependent. Negative error numbers are reserved by the
SCPI standard..
Return values:
<Error> List of:
<Error/event_number>,"<Error/event_description>[;<Device-
dependent info>]"
If the queue is empty, the response is 0,“No error“
Usage: Query only
SCPI confirmed
SYSTem:ERRor:ALL?
Queries the error/event queue for all unread items and removes them from the queue.
The response is a comma separated list of error number and a short description of the
error in FIFO order. Positive error numbers are instrument-dependent. Negative error
numbers are reserved by the SCPI standard.
Return values:
<Error> List of:
<Error/event_number>,"<Error/event_description>[;<Device-
dependent info>]"
If the queue is empty, the response is 0,“No error“
Usage: Query only
SCPI confirmed
SYST:PRESet
Resets the instrument to the default state, has the same effect as *RST.
Usage:
Event
SYSTem:BEEPer[:IMMediate]
Generates an immediate beep.
Usage:
Event
General Instrument Setup