Trigger
R&S
®
RTC1000
50User Manual 1335.7352.02 ─ 02
AUTO LEVEL
Automatic filter setting (default setting).
AC
The trigger signal is coupled through a 5 Hz high-pass filter which suppresses the DC
portion of the trigger signal. With a changing DC portion, the trigger level remains at
the set point in the AC signal.
DC
The trigger signal is coupled to the trigger circuit with all signal portions (AC and DC
voltage).
HF
The trigger signal is coupled through a high-pass filter with a minimum cut-off fre-
quency (-3 dB) of 30 kHz. Use this coupling type only with high frequency signals.
LOWPASS
The trigger signal is coupled through a low-pass filter with an upper cut-off frequency of
5 kHz. This filter removes higher frequencies and is available with AC and DC cou-
pling.
NOISE REJECT
A low-pass filter with a cut-off frequency of 100 MHz improves the noise performance
for the trigger amplifier. This filter removes higher frequencies and is available with AC
and DC coupling.
Note: The coupling types "LOWPASS" and "NOISE REJECT" cannot be activated
simultaneously.
Edge Trigger

Table of Contents

Other manuals for R&S RTC1000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S RTC1000 and is the answer not in the manual?

R&S RTC1000 Specifications

General IconGeneral
BrandR&S
ModelRTC1000
CategoryTest Equipment
LanguageEnglish

Related product manuals