EasyManuals Logo
Home>Rigol>Test Equipment>MSO2000A Series

Rigol MSO2000A Series Programming Guide

Rigol MSO2000A Series
376 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
RIGOL 2 Command System
2-90 MSO2000A/DS2000A Programming Guide
:DISPlay:GRADing:TIME
Syntax
:DISPlay:GRADing:TIME <time>
:DISPlay:GRADing:TIME?
Description
Set or query the persistence time and the unit is s.
Parameter
Name Type Range Default
<time> Discrete {MIN|0.05|0.1|0.2|0.5|1|2|5|10|20|INFinite} MIN
Explanation
MIN: set the persistence time to its minimum to view the waveform changing in high refresh rate.
Specific Values: a certain value between 0.05s and 20s, enable to observe glitch that changes
relatively slowly or glitch with low occurrence probability.
INFinite: in this mode, the oscilloscope displays the newly acquired waveform without clearing the
waveform formerly acquired. Enable to measure noise and jitter as well as capture incidental events.
Return Format
The query returns the persistence time set.
:DISPlay:WBRightness
Syntax
:DISPlay:WBRightness <time>
:DISPlay:WBRightness?
Description
Set or query the waveform brightness and the unit is %.
Parameter
Name Type Range Default
<time> Integer 0 to 100 50
Return Format
The query returns an integer between 0 and 100.

Table of Contents

Other manuals for Rigol MSO2000A Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol MSO2000A Series and is the answer not in the manual?

Rigol MSO2000A Series Specifications

General IconGeneral
BrandRigol
ModelMSO2000A Series
CategoryTest Equipment
LanguageEnglish

Related product manuals