RIGOL 2 Command System
2-338 MSO2000A/DS2000A Programming Guide
:WAVeform:STARt
Syntax
:WAVeform:STARt <sta>
:WAVeform:STARt?
Description
Set or query the start position of internal memory waveform reading.
Parameter
Name Type Range Default
<sta> Integer
NORMal: 1 to 1400
MAX: 1 to the number of effective point currently
on the screen
RAW: 11 to the current maximum memory depth
1
Explanation
For the memory depth, refer to the
:ACQuire:MDEPth command.
The range of the start position is limited by the current waveform reading mode (refer to
the
:WAVeform:MODE command).
Return Format
The query returns an integer.