RIGOL 2 Command System
2-256 MSO2000A/DS2000A Programming Guide
:TIMebase[:MAIN]:SCALe
Syntax
:TIMebase[:MAIN]:SCALe <scale_value>
:TIMebase[:MAIN]:SCALe?
Description
Set or query the main time base and the unit is s/div.
Parameter
Name Type Range Default
<scale_value> Real
The range differs for different mode
[1]
:
Normal: 1ns
[2]
to 1ks
ROLL: 200ms to 1ks
1us/div
Note
[1]
: Refer to the :TIMebase:MODE command.
Note
[2]
: This value is different for different model. For MSO2202A/DS2202A/MSO2202A-S/DS2202A-S, it is 2ns. For
MSO2102A/DS2102A/MSO2102A-S/DS2102A-S and MSO2072A/DS2072A/MSO2072A-S/DS2072A-S, it is 5ns.
Return Format
The query returns the current main time base in scientific notation.