2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-127
:MASK Commands
:MASK:ENABle
:MASK:SOURce
:MASK:OPERate
:MASK:MDISplay
:MASK:SOOutput
:MASK:OUTPut
:MASK:X
:MASK:Y
:MASK:CREate
:MASK:PASSed?
:MASK:FAILed?
:MASK:TOTal?
:MASK:RESet
:MASK:DATA
:MASK:ENABle
Syntax
:MASK:ENABle <bool>
:MASK:ENABle?
Description
Enable or disable the pass/fail test function; or query the current status of the pass/fail test function.
Parameter
Name Type Range Default
<bool> Bool {{0|OFF}|{1|ON}} 0|OFF
Explanation
In the following states, the pass/fail test function cannot be enabled and this command is invalid.
In the X-Y time base mode
In the ROLL time base mode
In the Y-T time base mode, when the horizontal time base is set to 200ms/div or slower and the
instrument enters the slow sweep mode
When the waveform recording is enabled
Return Format
The query returns 0 or 1.