RIGOL 2 Command System
2-248 MSO2000A/DS2000A Programming Guide
:SYSTem:OPTion:VALid?
Syntax
:SYSTem:OPTion:VALid? <module>
Description
Query the status of the options of the oscilloscope.
Parameter
Name Type Range Default
<module> Discrete {TRIGger|DECode| CAN|MEMDepth} --
Explanation
TRIGger: advanced trigger option
DECode: decoding option
CAN: CAN protocol analysis option (include CAN trigger and CAN decoding)
MEMDepth: deep memory depth option
Return Format
The query returns 0, 1, 2 or 3.
0: the option is not installed
1: the trial period of the option expires
2: original option is installed
3: valid trial option is installed
:SYSTem:PON
Syntax
:SYSTem:PON <power_on>
:SYSTem:PON?
Description
Set or query the configuration to be recalled when the oscilloscope is powered on after power-off.
Parameter
Name Type Range Default
<power_on> Discrete {LATest|DEFault} DEFault
Explanation
LATest: restore the oscilloscope to the setting at the last power-off.
DEFault: restore the oscilloscope to the factory setting.
Return Format
The query returns LAT or DEF.