2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-255
:TIMebase[:MAIN]:OFFSet
Syntax
:TIMebase[:MAIN]:OFFSet <offset>
:TIMebase[:MAIN]:OFFSet?
Description
Set or query the horizontal position of the main time base and the unit is s.
Parameter
Name Type Range Default
<offset> Real
RUN:
When TimeScale is lower than 10ms:
-MemDepth/SamplingRate to 1s
TimeScale is greater than or equal to 10ms:
-MemDepth/SamplingRate to 100×TimeScale
STOP: -7000s to 7000s
ROLL RUN: not available
ROLL STOP: -7000s to 0
0
Note:
For the MemDepth, refer to the
:ACQuire:MDEPth command.
For the SamplingRate, refer to the
:ACQuire:SRATe? command.
For the TimeScale, refer to the
:TIMebase[:MAIN]:SCALe command.
Return Format
The query returns the horizontal position in the scientific notation.