2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-221
:SAVE:MTESt[:STARt]
Syntax
:SAVE:MTESt[:STARt] {<internal_loc>|<file_spec>}
Description
Store the current pass/fail test mask into the internal or external memory in “*.pf” format.
Parameter
Name Type Range Default
<internal_loc> ASCII string C:\LocalPF0 to C:\LocalPF9 --
<file_spec> ASCII string Refer to the Explanation --
Explanation
This command is only available when the pass/fail test function is enabled (refer to
:MASK:ENABle).
In internal storage, if the specified location already contains a file, the original file will be overwritten.
In external storage, <file_ spec> includes the external storage directory and the filename with suffix.
Its length cannot exceeds 64 characters. Wherein, the storage directory can only be D:\. If the
specified directory already contains a file with the same filename, the original file will be overwritten.
Example
:SAVE:MTESt C:\LocalPF3.pf /*Store the current pass/fail test mask into the internal memory*/

Table of Contents

Other manuals for Rigol MSO2000A Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol MSO2000A Series and is the answer not in the manual?

Rigol MSO2000A Series Specifications

General IconGeneral
BrandRigol
ModelMSO2000A Series
CategoryTest Equipment
LanguageEnglish

Related product manuals