[:TRACe<n>]:DATA:POINts:INTerpolate
Syntax
[:TRACe<n>]:DATA:POINts:INTerpolate <mode>
[:TRACe<n>]:DATA:POINts:INTerpolate?
Description
Set or query the interpolation mode between the defined points of the arbitrary waveform of the built-in
signal source.
Parameter
Name Type Range Default
<mode> Discrete {LINear|OFF} OFF
<n> Discrete {1|2} 1
Explanation
LINear: linear interpolation. The waveform editor connects two defined points using a straight line.
OFF: turn off the interpolation mode. The waveform editor will keep a constant voltage level between
two points and create a ladder-like waveform.
Return Format
The query returns LIN or OFF.