2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
:MEASure:OVERshoot:SMAXimum?
:MEASure:OVERshoot:SMINimum?
:MEASure:OVERshoot:SCURrent?
:MEASure:OVERshoot:SAVerage?
:MEASure:OVERshoot:SDEViation?
Syntax 1
:MEASure:OVERshoot:SMAXimum <chan>
:MEASure:OVERshoot:SMINimum <chan>
:MEASure:OVERshoot:SCURrent <chan>
:MEASure:OVERshoot:SAVerage <chan>
:MEASure:OVERshoot:SDEViation <chan>
Description 1
Enable the overshoot measurement function of the specified channel and enable the statistic function.
Syntax 2
:MEASure:OVERshoot:SMAXimum? <chan>
:MEASure:OVERshoot:SMINimum? <chan>
:MEASure:OVERshoot:SCURrent? <chan>
:MEASure:OVERshoot:SAVerage? <chan>
:MEASure:OVERshoot:SDEViation? <chan>
Description 2
Enable the overshoot measurement function of the specified channel; enable the statistic function; return
the maximum value, minimum value, current value, average value or standard deviation.
Parameter
Name Type Range Default
<chan> Discrete {CHANnel1|CHANnel2|MATH} --
Return Format
The query returns the measurement result in scientific notation.