Common Measurement Settings
R&S
®
FSW
345User Manual 1173.9411.02 ─ 19
Source State............................................................................................................... 345
Source Power..............................................................................................................345
Source Offset.............................................................................................................. 345
Source Frequency Coupling........................................................................................346
(Manual) Source Frequency........................................................................................346
(Automatic) Source Frequency (Numerator/Denominator/Offset)...............................346
Result Frequency Start............................................................................................... 346
Result Frequency Stop................................................................................................347
Source State
Activates or deactivates control of an external generator.
Remote command:
SOURce:EXTernal[:STATe] on page 886
Source Power
The output power of the external generator. The default output power is -20 dBm. The
range is specified in the data sheet.
Remote command:
SOURce:EXTernal:POWer[:LEVel] on page 886
Source Offset
Constant level offset for the external generator. Values from -200 dB to +200 dB in 1
dB steps are allowed. The default setting is 0 dB. Offsets are indicated by the "LVL"
label in the channel bar (see also "Displayed Information and Errors" on page 341).
With this offset, attenuators or amplifiers at the output connector of the external gener-
ator can be taken into account for the displayed output power values on screen or dur-
ing data entry, for example. Positive offsets apply to an amplifier and negative offsets
to an attenuator subsequent to the external generator.
Remote command:
SOURce:POWer[:LEVel][:IMMediate]:OFFSet on page 886