Common Analysis and Display Functions
R&S
®
FSW
495User Manual 1173.9411.02 ─ 19
Marker Position (X-value)
Defines the position (x-value) of the marker in the diagram.
Remote command:
CALCulate<n>:MARKer<m>:X on page 932
CALCulate<n>:DELTamarker<m>:X on page 931
Frame (Spectrogram only)
Spectrogram frame the marker is assigned to.
Remote command:
CALCulate<n>:MARKer<m>:SGRam|SPECtrogram:FRAMe on page 945
Marker Type
Toggles the marker type.
The type for marker 1 is always "Normal", the type for delta marker 1 is always "Delta".
These types cannot be changed.
Note: If normal marker 1 is the active marker, switching the "Mkr Type" activates an
additional delta marker 1. For any other marker, switching the marker type does not
activate an additional marker, it only switches the type of the selected marker.
"Normal"
A normal marker indicates the absolute value at the defined position
in the diagram.
"Delta"
A delta marker defines the value of the marker relative to the speci-
fied reference marker (marker 1 by default).
Remote command:
CALCulate<n>:MARKer<m>[:STATe] on page 932
CALCulate<n>:DELTamarker<m>[:STATe] on page 930
Reference Marker
Defines a marker as the reference marker which is used to determine relative analysis
results (delta marker values).
If a fixed reference point is configured (see "Defining a Fixed Reference"
on page 497), the reference point ("FXD") can also be selected instead of another
marker.
Remote command:
CALCulate<n>:DELTamarker<m>:MREF on page 930
Linking to Another Marker
Links the current marker to the marker selected from the list of active markers. If the x-
axis value of the inital marker is changed, the linked marker follows on the same x-
position. Linking is off by default.
Using this function you can set two markers on different traces to measure the differ-
ence (e.g. between a max hold trace and a min hold trace or between a measurement
and a reference trace).
Remote command:
CALCulate<n>:MARKer<m1>:LINK:TO:MARKer<m2> on page 931
CALCulate<n>:DELTamarker<m1>:LINK:TO:MARKer<m2> on page 929
CALCulate<n>:DELTamarker<m>:LINK on page 929
Marker Usage