Remote Commands
R&S
®
FSW
932User Manual 1173.9411.02 ─ 19
Example:
CALC:MARK4:LINK:TO:MARK2 ON
Links marker 4 to marker 2.
Manual operation: See "Linking to Another Marker" on page 300
CALCulate<n>:MARKer<m>[:STATe] <State>
This command turns markers on and off. If the corresponding marker number is cur-
rently active as a deltamarker, it is turned into a normal marker.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK3 ON
Switches on marker 3.
Manual operation: See "Marker State" on page 299
See "Marker Type" on page 300
CALCulate<n>:MARKer<m>:TRACe <Trace>
This command selects the trace the marker is positioned on.
Note that the corresponding trace must have a trace mode other than "Blank".
If necessary, the command activates the marker first.
Parameters:
<Trace> 1 to 6
Trace number the marker is assigned to.
Example:
CALC:MARK3:TRAC 2
Assigns marker 3 to trace 2.
Manual operation: See "Assigning the Marker to a Trace" on page 300
CALCulate<n>:MARKer<m>:X <Position>
This command moves a marker to a particular coordinate on the x-axis.
If necessary, the command activates the marker.
If the marker has been used as a delta marker, the command turns it into a normal
marker.
Parameters:
<Position> Numeric value that defines the marker position on the x-axis.
The unit is either Hz (frequency domain) or s (time domain) or
dB (statistics).
Range: The range depends on the current x-axis range.
Example:
CALC:MARK2:X 1.7MHz
Positions marker 2 to frequency 1.7 MHz.
Analyzing Measurements (Basics)