Remote Commands
R&S
®
FSW
931User Manual 1173.9411.02 ─ 19
Parameters:
<Trace> Trace number the marker is assigned to.
Example:
CALC:DELT2:TRAC 2
Positions delta marker 2 on trace 2.
CALCulate<n>:DELTamarker<m>:X <Position>
This command moves a delta marker to a particular coordinate on the x-axis.
If necessary, the command activates the delta marker and positions a reference
marker to the peak power.
Parameters:
<Position> Numeric value that defines the marker position on the x-axis.
The position is relative to the reference marker.
To select an absolute position you have to change the delta
marker mode with CALCulate<n>:DELTamarker:MODE
on page 929.
A query returns the absolute position of the delta marker.
Range: The value range and unit depend on the measure-
ment and scale of the x-axis.
Example:
CALC:DELT:X?
Outputs the absolute x-value of delta marker 1.
Manual operation: See "Marker 1/2/3/4" on page 282
See "Marker 1/2/3" on page 287
See "Marker Position (X-value)" on page 299
CALCulate<n>:MARKer<m>:AOFF
This command turns all markers off.
Example:
CALC:MARK:AOFF
Switches off all markers.
Usage: Event
Manual operation: See "All Markers Off" on page 496
CALCulate<n>:MARKer<m1>:LINK:TO:MARKer<m2> <State>
This command links normal marker <m1> to any active normal marker <m2>.
If you change the horizontal position of marker <m2>, marker <m1> changes its hori-
zontal position to the same value.
Parameters:
<State> ON | OFF
*RST: OFF
Analyzing Measurements (Basics)