929User Manual 1173.9411.02 ─ 19
CALCulate<n>:DELTamarker:AOFF
This command turns all delta markers off.
Example:
CALC:DELT:AOFF
Turns all delta markers off.
Usage: Event
CALCulate<n>:DELTamarker<m>:LINK <State>
This command links delta marker <m> to marker 1.
If you change the horizontal position (x-value) of marker 1, delta marker <m> changes
its horizontal position to the same value.
Tip: to link any marker to a different marker than marker 1, use the CALCulate<n>:
DELTamarker<m1>:LINK:TO:MARKer<m2> or CALCulate<n>:MARKer<m1>:
LINK:TO:MARKer<m2> commands.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:DELT2:LINK ON
Manual operation: See "Linking to Another Marker" on page 300
CALCulate<n>:DELTamarker<m1>:LINK:TO:MARKer<m2> <State>
This command links delta marker <m1> to any active normal marker <m2>.
If you change the horizontal position of marker <m2>, delta marker <m1> changes its
horizontal position to the same value.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:DELT4:LINK:TO:MARK2 ON
Links the delta marker 4 to the marker 2.
Manual operation: See "Linking to Another Marker" on page 300
CALCulate<n>:DELTamarker:MODE <Mode>
This command defines whether the position of a delta marker is provided as an abso-
lute value or relative to a reference marker. Note that when the position of a delta
marker is queried, the result is always an absolute value (see CALCulate<n>:
DELTamarker<m>:X on page 931)!