Common Analysis and Display Functions
R&S
®
FSW
516User Manual 1173.9411.02 ─ 19
This function is not available for Spectrum Emission Mask measurements or measure-
ments on I/Q-based data.
For details see chapter 7.4.1.11, "Demodulating Marker Values and Providing Audio
Output", on page 492.
Marker Demodulation State........................................................................................ 516
Continuous Demodulation...........................................................................................516
Marker Stop Time........................................................................................................516
Modulation...................................................................................................................517
Squelch....................................................................................................................... 517
Squelch level...............................................................................................................517
Marker Demodulation State
Activates or deactivates the demodulation output.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation[:STATe] on page 966
Continuous Demodulation
If activated, the signal is demodulated continuously (not only at the marker position)
and sent to the audio output. This allows you to monitor the frequency range acousti-
cally (assuming the sweep time is long enough).
For zero span and EMI measurements, demodulation is always active continuously.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:CONTinuous
on page 965
Marker Stop Time
Defines how long the sweep is stopped at the marker position to output the demodula-
ted signal.
For zero span measurements, demodulation is always active continuously, regardless
of the marker stop time.