Remote Commands
R&S
®
FSW
965User Manual 1173.9411.02 ─ 19
11.8.3.14 Marker Demodulation
The following commands control the demodulation of AM and FM signals at the marker
position.
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:CONTinuous................................... 965
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:HOLDoff..........................................965
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:SELect............................................965
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation[:STATe].......................................... 966
[SENSe:]DEMod:SQUelch:LEVel.....................................................................................966
[SENSe:]DEMod:SQUelch[:STATe]................................................................................. 966
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:CONTinuous <State>
This command turns continuous demodulation of the signal at the marker position in
the frequency domain on and off.
In the time domain continuous demodulation is always on.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC2:MARK3:FUNC:DEM:CONT ON
Switches on the continuous demodulation.
Manual operation: See "Continuous Demodulation" on page 516
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:HOLDoff <Duration>
This command defines for how long the the signal at the marker position is demodula-
ted.
In the time domain continuous demodulation is always on.
Parameters:
<Duration> Range: 10 ms to 1000 s
*RST: Marker demodulation = OFF
Example:
CALC:MARK:FUNC:DEM:HOLD 3s
Manual operation: See "Marker Stop Time" on page 516
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:SELect <DemodMode>
This command selects the demodulation mode at the marker position.
Parameters:
<DemodMode> AM
AM demodulation
FM
FM demodulation
*RST: AM
Analyzing Measurements (Basics)