EasyManua.ls Logo

Rohde & Schwarz FSW26 - Page 981

Rohde & Schwarz FSW26
1150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSW
981User Manual 1173.9411.02 ─ 19
Parameters:
<Offset> Numeric value.
*RST: 0
Default unit: dB
Manual operation: See "Y-Offset" on page 531
CALCulate:LIMit<k>:LOWer:SHIFt <Distance>
This command moves a complete lower limit line vertically.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Parameters:
<Distance> Defines the distance that the limit line moves.
The unit depends on CALCulate:LIMit<k>:UNIT
on page 982.
Manual operation: See "Shift y" on page 534
CALCulate:LIMit<k>:LOWer:SPACing <InterpolType>
This command selects linear or logarithmic interpolation for the calculation of a lower
limit line from one horizontal point to the next.
Parameters:
<InterpolType> LINear | LOGarithmic
*RST: LIN
Manual operation: See "X-Axis" on page 533
See "Y-Axis" on page 533
CALCulate:LIMit<k>:LOWer:STATe <State>
This command turns a lower limit line on and off.
Before you can use the command, you have to select a limit line with CALCulate:
LIMit<k>:NAME on page 982.
Parameters:
<State> ON | OFF
*RST: OFF
Usage: SCPI confirmed
Manual operation: See "Visibility" on page 530
CALCulate:LIMit<k>:LOWer:THReshold <Threshold>
This command defines a threshold for relative limit lines.