Remote Commands
R&S
®
FSW
983User Manual 1173.9411.02 ─ 19
Manual operation: See "Data points" on page 533
CALCulate:LIMit<k>:UPPer:MARGin <Margin>
This command defines an area around an upper limit line where limit check violations
are still tolerated.
Parameters:
<Margin> numeric value
*RST: 0
Default unit: dB
Manual operation: See "Margin" on page 533
CALCulate:
LIMit<k>:UPPer:MODE <Mode>
This command selects the vertical limit line scaling.
Parameters:
<Mode> ABSolute
Limit line is defined by absolute physical values.
The unit is variable.
RELative
Limit line is defined by relative values related to the reference
level (dB).
*RST: ABSolute
Manual operation: See "X-Axis" on page 533
CALCulate:LIMit<k>:UPPer:OFFSet <Offset>
This command defines an offset for a complete upper limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Parameters:
<Offset> Numeric value.
*RST: 0
Default unit: dB
Manual operation: See "Y-Offset" on page 531
CALCulate:LIMit<k>:UPPer:SHIFt <Distance>
This command moves a complete upper limit line vertically.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Analyzing Measurements (Basics)