PULSE/VIDEO Output R&S®SML / R&S®SMV03
1090.3123.12 E-6 4.48
Trigger Mode
Sel
ection of trigger mode.
Auto The pulse generator is triggered automatically. The pulse period is
as entered under Pulse Period.
Single The pulse generator is triggered manually. The pulse period is
determined by the user.
Ext Single The pulse generator is externally triggered. The pulse
period is determined by an external signal at the PULSE input.
Ext Gated The pulse generator is triggered if the gate signal is active.
IEC/IEEE-bus command :TRIG:PULS:SOUR AUTO
Execute Single Pulse
Starts a single pulse. This function is displayed and is effective only if
Single Mode is selected.
IEC/IEEE-bus command :TRIG:PULS:IMM
Ext Trigger Slope
Selection of active edge of external trigger signal.
Pos The pulse generator is triggered on the positive edge of the
external signal.
Neg The pulse generator is triggered on the negative edge of the
external signal.
IEC/IEEE-bus command :TRIG:PULS:SLOP POS
Ext Gated Input Polarity
Definition of active level of gate signal (HIGH or LOW). Normal (HIGH) and
Inverse (LOW) are available.
IEC/IEEE-bus command :TRIG:PULS:EGAT:POL NORM