EasyManua.ls Logo

Rohde & Schwarz SML01 - Reference Frequency

Rohde & Schwarz SML01
299 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Test sequence R&S®SML / R&S®SMV03
1090.3123.12 10.8 E-6
Reference frequency
Internal reference output
Note: Let the R&S SML / R&S SMV03 warm up for at least 2 hours before starting a measurement.
Test setup Connect test receiver (spectrum analyzer mode) to REF EXT output
on rear panel of R&S SML / R&S SMV03.
Settings on the test receiver - Center frequency = 10 MHz
- S
pan = 20 Hz
- Resolution bandwidth = 10 Hz
Measurement Set marker on peak and read off frequency.
Analysis The frequency deviation must not exceed the sum of the deviations
resulting from the frequency error in the rated temperature range and
from aging.
External reference input
Test setup Connect test receiver (spectrum analyzer mode) to RF connector on
R&S SML / R&S SMV03.
Connect signal generator to REF IN connector on rear panel of R&S
SML / R&S SMV03.
Connect references of test receiver and signal generator
Settings on the test receiver - C
enter frequency = test frequency of R&S SML / R&S SMV03
- S
pan = 100 Hz
- Resolution bandwidth = 10 Hz
Settings on R&S SML /
R&S SMV03
- Switch reference oscillator to external
Settings on the signal
generator
- Frequency = 10 MHz
- Level = 7 dBm
Measurement Set marker on peak and read off frequency
Analysis The frequency deviation must not exceed the sum of the deviations
resulting from the frequency error in the rated temperature range and
from aging.

Table of Contents

Related product manuals