R&S®SML / R&S®SMV03 Test sequence
1090.3123.12 10.7 E-6
- FREQ/CENTER/STOP FREQUENCY,
- FM DEMOD ON,
- Demodulation bandwidth 100 kHz,
- Span 200 Hz,
- MEAS TIME 10 ms.
Settings on R&S SML /
R&S SMV03
- LEVEL: 0 dBm
- UTILITIES REF OSC SOURCE EXTERN
Measurement Setting on spectrum analyzer: TRIGGER EXTERN
- external triggering on positive edge at 1.4 V
Send target frequency from process controller.
The build-up trace appears on the screen of the externally
triggered spectrum analyzer. The settling time can be determined
with the aid of the cursor.
Repeat measurement with start and target frequencies interchanged.
The following hops need to be measured:
Table 10-3 Measuring the frequency setting time
Start frequency Target frequency
806 MHz 808 MHz
606 MHz 1075 MHz
1075 MHz 76 MHz
76 MHz 1075 MHz
1210.5 MHz 2200.000001 MHz
2800 MHz 1818 MHz