Test sequence R&S®SML / R&S®SMV03
1090.3123.12 10.26 E-6
Setting on the modulation
analyzer
- Stereo signal as for crosstalk attenuation measurement
- Demodulation: FM STEREO
- CHANNEL: L or R
- FILTER: CCIR WT or UNWT
- DETECTOR RMS
- Relative deviation measurement
- DEEMPHASIS 50 µs
Measurement
Switch on the left or right channel on the stereo coder and carry
out a relative measurement. Then switch off the useful signal on
the stereo coder and read off the S/N ratio. Carry out this
measurement for both filters (weighted and unweighted). Then
switch on the right channel and repeat the same measurement.
Phase modulation
PhiM deviation setting
Test setup Test setup 1, demodulator mode
Settings on R&S SML /
R&S SMV03
- LEVEL: 0 dBm
- FREQ: 1 GHz
- MODULATION PhiM PhiM DEVIATION: 5 rad
PhiM SOURCE: LFGen
LFGenFreq = 1 kHz
Settings on the test receiver - D
emodulation: PhiM
- D
etector: Peak
- Filter: 20 Hz to 23 kHz
(for the moment measurement must still be carried out with 50 Hz /
100 kHz, since filters are not yet implemented)
Measurement Read off the phase deviation from the test receiver
Note: Since the R&S SML / R&S SMV03 has a purely digital deviation generator, it is adequate
to test its functionality by taking a measurement at one deviation setting and one frequency
only.