EasyManua.ls Logo

Rohde & Schwarz SML01 - Source2 System; Source2:Frequency Subsystem

Rohde & Schwarz SML01
299 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S®SML / R&S®SMV03 SOURce2:FREQuency
1090.3123.12 E-6 6.41
SOURce2 System
The SOURce2 system contains the commands to configure the LF signal source. The LF signal source
s designated as INT if it is used as a modulation source, if it is used as an LF generator, it is designated
as SOURce2.
The commands to set the output voltage of the LF generator are in the OUTPut2 system.
Subsystems Settings
:SOURce2
:FREQuency
:SWEep
Frequency with CW and sweep operation
LF sweep
SOURce2:FREQuency Subsystem
This subsystem contains the commands for the frequency settings including the sweeps.
Command Parameters Default
Unit
Remark
:SOURce2
:FREQuency
[:CW|:FIXed]
:MANual
:MODE
:STARt
:STOP
0.1 Hz to 1 MHz
0.1 Hz to 1 MHz
CW | FIXed | SWEep
0.1 Hz to 1 MHz
0.1 Hz to 1 MHz
Hz
Hz
Hz
Hz
:SOURce2:FREQuency[:CW | :FIXed] 0.1 Hz to 1 MHz
The command sets the frequency for the CW mode.
Example: :SOUR2:FREQ:CW 1kHz RST value is 1 kHz
:SOURce2:FREQuency:MANual 0.1 Hz to 1 MHz
The command sets the frequency if SOUR2:SWE:MODE MAN and SOUR2:FREQ:MODE SWE are
set. In this case, only frequency values between the settings :SOUR2:FREQ:STAR and to:STOP
are allowed.
Example: :SOUR2:FREQ:MAN 1kHz *RST value is 1 kHz

Table of Contents