TRIGger R&S®SML / R&S®SMV03
1090.3123.12 E-6 6.52
UNIT System
Using this command, the basic unit of the RF level only can be changed and set. Simultaneously, the
displayed unit is changed accordingly.
Command Parameters
Default
Unit
Remark
:UNIT
:POWer
DBM | VOLT | DBUV | V
:UNIT:POWer DBM | VOLT | DBUV | V
Using this command, the basic unit of the RF level only can be changed and set. Simultaneously,
the displayed unit is changed accordingly.
Note: The LF generator level is always in V!
Example: after *RST :
:pow 0
:unit:pow? Answer : DBM
:pow? Answer : 0.000000E+00
:unit:pow dbuv
:unit:pow? Answer : DBUV
:pow? Answer : 1.069897E+02
:pow 0
:unit:pow? Answer : DBUV
:pow? Answer : 0.000000E+00
:unit:pow dbm
:unit:pow? Answer : DBUV
:pow? Answer : -1.069897E+02