Switchover to Remote Control R&S®SML / R&S®SMV03
1090.3123.12 E-6 5.4
Return to Manual Operation
Return to manual operation can be made via the front panel or the IEC/IEEE bus.
Manually: Press [LOCAL] key.
Note:
Before switchover, command processing must be completed as otherwise
switchover to remote control is effected immediately.
The [LOCAL] key can be disabled by the universal command LLO in order to
prevent unintentional switchover. In this case, switchover to manual control is
only possible via the IEC/IEEE bus.
The [LOCAL] key can be enabled again by deactivating the REN control line
of the IEC/IEEE bus.
Via IEC/IEEE bus: ...
CALL IBLOC(generator%) Set instrument to manual control
...
Remote Control via RS-232-C Interface
Setting the Transmission Parameters
To enable error-free and correct data transmission, the parameters of the instrument and the controller
should have the same setting. To prevent any problems during binary data transmission, the RS-232-C
interface should be set to 8 data bits, “No parity” and 1 stop bit. This data format corresponds to the
IEEE P1174 draft standard.
The baud rate and handshake can be manually changed in the Utilities - System - RS232 menu.
Call Utilities – System - RS232 menu.
Select desired baud rate and handshake.
Terminate input using the [1x/ENTER] key.
Indications during Remote Control
The remote control state is indicated by "Remote" in the STATUS line. In the REMOTE state, the
STATUS page is always displayed.
Return to Manual Operation
Return to manual operation can be made via the front panel.
Press [LOCAL] key.
Note: Before switchover, command processing must be completed as otherwise
switchover to remote control is effected immediately.