Test sequence  R&S®SML / R&S®SMV03 
1090.3123.12 10.18  E-6 
Overvoltage protection 
Measuring equipment  -  Sinewave generator (table "Measuring Equipment and Accessories", item 9)
Test setup   Conn
ect the sinewave generator to the RF output of the R&S SML / 
R&S SMV03. 
Settings on R&S SML / 
R&S SMV03 
- FREQ: 
100 MHz 
- LEVEL: 
- 140 dBm 
 
Settings on the sinewave 
generator 
- FREQ: 
20 kHz 
- Output impedance = 50 Ohm 
- Level = 1 V 
- Level offset = ± 5 V
Measurement   Increase the output level of the sinewave generator to a maximum of 10 
V (EMF) 
 The overvoltage protection must respond to a voltage (offset + 
EMF/2) > 3.7 V and < 7.5 V for both polarities. 
 
Internal modulation generator 
Note:  The setting time refers purely to computer time and therefore does not need to be 
measured. 
 
Level accuracy 
Measuring equipment  -  AC voltmeter (table "Measuring Equipment and Accessories", item 10)
Test setup   Conn
ect AC voltmeter to LF connector of R&S SML / R&S SMV03  
Settings on R&S SML / 
R&S SMV03 
- LF OUTPUT STATE ON 
- LF OUTPUT LFGen 1 kHz 
- LF OUTPUT VOLTAGE: 
1 mV, 10 mV, 100 mV, 1 V, 4 V 
 
Measurement   Measure output level 
 
Frequency response 
Measuring equipment  -  AC voltmeter (table "Measuring Equipment and Accessories", item 10) 
Test setup   Connect AC voltmeter to LF connector of R&S SML / R&S SMV03. 
 
Settings on R&S SML / 
R&S SMV03 
- LF OUTPUT STATE ON 
- LF OUTPUT VOLTAGE: 
1 V, 4 V 
- LF OUTPUT LFGen: 
10 Hz to 500 kHz, 5 reference values / decade 
 
Measurement   Measure the frequency response  
 The frequency response is the difference between the highest and 
lowest level.