DISPlay R&S®SML / R&S®SMV03
1090.3123.12 E-6 6.10
DISPLAY System
This system contains the commands to configure the screen. If system security is activated using
command SYSTem:SECurity ON, the display cannot be switched on and off arbitrarily (cf. below).
Command Parameter Default
Unit
Remark
:DISPlay
:ANNotation
[:ALL]
:AMPLitude
:FREQuency
ON | OFF
ON | OFF
ON | OFF
:DISPlay:ANNotation
The commands determining whether frequency and amplitude are indicated are under this node.
Caution: With SYSTem:SECurity ON, the indications cannot be switched from OFF to ON. In
this case *RST does not influence the ANNotation settings either. With
SYSTem:SECurity OFF, the *RST value is ON for all ANNotation parameters.
:DISPlay:ANNotation[:ALL] ON | OFF
The command switches the frequency and amplitude indication on or off.
Command :DISP:ANN:ALL ON can only be executed if SYST:SEC is set to OFF.
Example: :DISP:ANN:ALL ON With SYST:SEC OFF - *RST value is ON
:DISPlay:ANNotation:AMPLitude ON | OFF
The command switches on or off the amplitude indication.
Command :DISP:ANN:AMPL ON can only be executed if SYST:SEC is set to OFF.
Example: :DISP:ANN:AMPL ON With SYST:SEC OFF - *RST value is ON
:DISPlay:ANNotation:FREQuency ON | OFF
The command switches on or off the frequency indication.
Command :DISP:ANN:FREQ ON can only be executed if SYST:SEC is set to OFF.
Example: :DISP:ANN:FREQ ON With SYST:SEC OFF - *RST value is ON