R&S®SML / R&S®SMV03  MEMory / OUTPut 
1090.3123.12   E-6 6.11
MEMory System 
This system contains the commands for the memory management of the R&S SML / R&S SMV03. 
 
Command Parameter  Default 
Unit 
Remark 
:MEMory 
 :NSTates? 
 
Query only 
:MEMory:NSTates? 
The command returns the number of *SAV/*RCL memories available. The R&S SML / R&S SMV03 
has 50 *SAV/*RCL memories in total.  
Example:  :MEM:NST? Response: 50 
OUTPut System 
This system contains the commands specifying the characteristics of the RF, LF and Pulse output 
sockets. The following numbers are assigned to these outputs: 
OUTPut1:  RF output, 
OUTPut2:  LF output, 
OUTPut3:  PULSE/VIDEO output. 
 
Command Parameter  Default 
Unit 
Remark 
:OUTPut1|2|3 
  :AFIXed 
  :RANGe 
  :LOW? 
:AMODe 
 :POLarity 
  :PULSe 
 :SOURce 
 [:STATe] 
  :PON 
 :VOLTage 
AUTO | FIXed 
 
NORMal | INVerted 
OFF | PULSegen | VIDeo 
OFF | ON 
OFF | UNCHanged 
0 V...4 V 
 
V
:OUTPut1:AFIXed:RANGe:LOW? 
The command returns the (only querable) lower value of the non-interrupting level range in 
"Attenuator Mode Fixed". 
Example: :OUTP:AFIX:RANG:LOW?  
:OUTPut1:AMODe   AUTO | FIXed 
The command switches over the operating mode of the attenuator (A
ttenuator MODe) at the RF 
output (output1).  
AUTO   Normal setting. The electronically switched attenuator switches in steps of 5 dB at 
fixed points. 
FIXed   Level settings are made without switching the attenuator (see RF level section "Non-
Interrupting Level Setting"). 
Example:  :OUTP:AMOD AUTO  *RST value is AUTO