R&S®SML / R&S®SMV03  Front Panel 
1090.3123.12   E-6 1.5
Explanation of Front and Rear Panel 
Elements of the Front Panel 
3
MOD
LF
RF 50
I
Q
1147.7509.13
dB?V
?V
mV
VECTOR SIGNAL GENERATOR 9 kHz ... 3.3 GHz SMV 03
. .
MAX 25 W
REF PWR
ASSIGN PRESET LOCALERROR
LEVEL
FREQ
SAVE
RCL
BACK
SELECT
HELP
STATUS
MOD
ON/OFF
RF
MENU1
0
1
2 3
4
5
6
7
8 9
.
-
.
G
M
k
m
n
x
1
?
ENTER
MENU2
ON/OFF
10
2 4 5
6
7
1
3
1 ON/OFF SWITCH 
 
The On/Off switch switches the instrument on ("I") or off 
("O"). 
 Cf. Chapter 1, Section "Switching On/Off the 
Instrument". 
 
2 DISPLAY 
Cf. Chapter 3 for the design of the display. 
 
3
Parameter field 
Parameters RF frequency and RF level can be entered 
directly by means of the parameter keys, alternatively to 
menu operation. Further, complete instrument settings can 
be stored and called. 
FREQ  Opens the setting of the RF frequency via value 
input or variation by means of a rotary knob. The 
current menu is maintained. Return to the menu 
by means of the [BACK] or [SELECT] key. 
(Setting of the RF frequency also in the 
FREQUENCY menu). 
LEVEL  Opens the setting of the RF level via value input 
or variation by means of a rotary knob. The 
current menu is maintained. Return to the menu 
by means of the [BACK] or [SELECT] key. 
(Setting of the RF level also in the LEVEL menu). 
SAVE  Opens the storing of the current instrument 
setting. Memory selection is effected by entering a 
number (1 to 50) and is finished by means of the 
[x1/ENTER] key. 
RCL  Opens the calling of an instrument setting stored. 
Memory selection is effected by entering a 
number (1 to 50) and is finished by means of the 
[x1/ENTER] key. 
 Cf. Chapter 3, Sections "Use of [FREQ] and 
[LEVEL] Keys", "RF Frequency", "RF Level" and 
"
S
toring and Calling of Instrument Settings". 
 
Fig. 1-1  Front panel view