R&S®SML / R&S®SMV03 Basic Operating Steps
1090.3123.12 E-6 3.5
Use of [FREQ] and [LEVEL] Keys
RF frequency and RF level can be set without menu operation as well using direct keys [FREQ] and
[LEVEL].
[FREQ] / [LEVEL] keys Press [FREQ] or [LEVEL] key.
The frequency or the level indication in the header field of the display is
marked. The current menu at the display is maintained.
Alter the value via a value input or the rotary knob.
Press [BACK] or [SELECT] key.
The menu cursor wraps to the position marked last in the menu.
Use of [RF ON/OFF] and [MOD ON/OFF]
RF signal and modulation can be switched on/off without menu operation as well using keys
[RF ON/OFF] or [MOD ON/OFF] (cf. Sections "[RF ON/OFF] Key" and "[MOD ON/OFF] Key").
[RF ON/OFF] key Press [RF ON/OFF] key.
The RF output signal is switched on/off.
IEC/IEEE-bus short command: :OUTP:STAT ON
[MOD ON/OFF] key Press [MOD ON/OFF] key.
Previous modulation(s) is switched off/on.
The modulations have to be switched on or off in the respective
modulation submenus.
IEC/IEEE-bus short command: :MOD:STAT OFF
Changing Unit of Level
For the level, the unit of the value set can be changed without a new value input.
Change level unit Activate Level parameter.
- Press [LEVEL] key or
- set menu cursor in the level menu to the setting value of the Amplitude
parameter.
Press the unit key with with the desired level unit.
The level is indicated in the desired unit.