List of Error Messages R&S®SML / R&S®SMV03
1090.3123.12 E-6 9.6
R&S SML / R&S SMV03-Specific Error Messages
Device-dependent Error – device-specific error; sets bit 3 in the ESR register.
Error code Error text in the case of queue poll
Error explanation
110
Output unleveled
The level control loop is deactivated.
115
Level overrange
The level is above the limit value guaranteed.
116
Level underrange
The level is below the limit value guaranteed.
117
Dynamic level range exceeded
The difference between the maximal and minimal value of a level list is above 20 dBm. An exact level
setting is no longer guaranteed.
135
Pulse input signal missing
No pulse input signal available.
140
This modulation forces other modulations OFF
A modulation has been switched on which cannot be used at the same time as an already active
modulation. The previous modulation has been switched off.
161
Output protection tripped
The overvoltage protection has been activated.
If a fault occurs (input overloaded), the output is automatically switched off (RF OFF).
To eliminate the fault, simply switch to RF ON (after the overload has been removed)
(no further action required).
171
Oven cold
The reference oscillator has not yet reached its operating temperature.
174
Reference PPL unlocked
The level control loop is deactivated.
175
Main PPL unlocked
The level control loop is deactivated.
180
Calibration failed
Calibration could not be executed.
181
REF OSC calibration data not used because ADJUSTMENT STATE is ON
The reference-oscillator calibration data are not used as long as ADJUSTMENT STATE is activated.
200
Cannot access hardware
The data transmission to a module was unsuccessful.
201
Function not supported by this hardware revision
A later version of certain parts of the instrument is necessary to execute the function selected.
202
Diagnostic A/D converter failure
Diagnostic A/D converter has failed.
203
Stereocoder, firmware missing