R&S®SML / R&S®SMV03 List of Error Messages
1090.3123.12 E-6 9.5
Device Specific Error - sets bit 3 in the ESR register
Error code Error text with queue poll
Explanation of error
-310
System error
This error message suggests an error within the instrument. Please inform your R&S service center.
-311
Memory error
Error in instrument memory.
-313
Calibration memory lost
Loss of stored calibration data. The YFOM and ALC AMP calibration data can be restored by means of
internal routines (see chapter 4, section "Calibration").
-314
Save/recall memory lost
Loss of the nonvolatile data stored with the command *SAV?.
-315
Configuration memory lost
Loss of the nonvolatile configuration data stored by the instrument.
-330
Self-test failed
The self-test could not be executed.
-350
Queue overflow
This error code is entered into the error queue instead of the actual error code when the error queue is
full. The code indicates that an error has occurred but has not been accepted. The error queue can
accept 5 entries.
-360
Communication error
An error has occurred during the transmission or reception of data on the IEC/IEEE bus or via the RS-
232-C interface.
Query Error – error in data request; sets bit 2 in the ESR register
Error code Error text with queue poll
Explanation of error
-410
Query INTERRUPTED
The query was interrupted.
Example: After a query, the instrument receives new data before the response has been sent completely.
-420
Query UNTERMINATED
The query is incomplete.
Example: The instrument is addressed as a talker and receives incomplete data.
-430
Query DEADLOCKED
The query cannot be processed.
Example: The input and output buffers are full; the instrument cannot continue operating.