Test sequence R&S®SML / R&S®SMV03
1090.3123.12 10.12 E-6
determined. The result is then referenced to 1 Hz bandwidth. The result
of this calculation is the value of the broadband noise.
Measurement - Measure the carrier power (P
T
)
- Detune the center frequency by 5 MHz
- Measure the channel power (P
K
)
Analysis The value of the broadband noise is determined from:
=
][
10
log10
10
HzChBw
PBBNoise
K
P
T
Residual FM
Test setup Test setup 1, demodulator mode
Settings on R&S SML /
R&S SMV03
- LEVEL: 0 dBm
- FREQ:
1 MHz, 20 MHz, 76 MHz, 100 MHz, 300 MHz, 605 MHz, 750 MHz,
900 MHz, 1 100 MHz, 1 211 MHz,
1 700 MHz, 2 200 MHz, 3 300 MHz
Settings on the test receiver - R
eference level = 0 dBm
- C
enter frequency = FREQ (R&S SML)
- Demodulation: FM
- Detector: RMS
- Filter: 20 Hz to 23 kHz
(for the moment measurement must still be carried out with 50 Hz /
100 kHz, since filters are not yet implemented)
Measurement Read off the frequency deviation.
Residual AM
Test setup Test setup 1, demodulator mode
Settings on R&S SML /
R&S SMV03
- LEVEL: 0 dBm
- FREQ:
5 MHz, 10 MHz, 76 MHz, 76 MHz + 1 Hz, 151.3125001 MHz,
255.2500001 MHz, 605.25 MHz, 605.2500001 MHz, 700 MHz, 807
MHz,
807 MHz + 1 Hz, 960 MHz, 1076 MHz, 1076 MHz + 1 Hz, 1250 MHz,
1800 MHz, 2010 MHz, 2200 MHz, 2655 MHz, 2655 MHz + 1 Hz, 2800
MHz, 2800 MHz + 1 Hz, 3300 MHz
Settings on the test receiver - R
eference level = 0 dBm
- Center frequency = FREQ (R&S SML)
- Demodulation: AM
- Detector: RMS
- Filter: 20 Hz to 23 kHz
(for the moment measurement must still be carried out with 50 Hz /
100 kHz, since filters are not yet implemented)
Measurement Read off the residual.