R&S®SML / R&S®SMV03 Utilities
1090.3123.12 E-6 4.61
Passwords for Accessing Protected Functions (Protect)
Calibration and service functions are password-protected. To access these functions, passwords (6-digit
numbers) have to be entered and confirmed with the [ENTER] key. These functions are automatically
locked out on power-up of the instrument.
Password 1 deactivates the lock for the calibration of Main Loop, Level Preset, LFGen Level, Leve,
IF Filter, Harm Filter, Mult Filterl.
Password 2 deactivates the lock for the calibration of RefOsc.
Password 3 factory internal
Password 4 factory internal
Access to protected functions is possible in the Utilities - Protect menu.
Menu selection: Utilities – Protect
Fig. 4-30 Utilities - Protect menu (preset setting)
Lock Level
x
Activation/deactivation of lock.
On The lock is active.
IEC/IEEE-bus command :SYST:PROT1 ON
Off The entry of the password is enabled automatically.
After entering the password, a pop-up menu is
displayed. The lock can be deactivated by selection Off.
IEC/IEEE-bus command :
S
YST:PROT1 OFF, 123456