R&S®SML / R&S®SMV03 SOURce2:SWEep
1090.3123.12 E-6 6.43
SOURce2:SWEep Subsystem
This subsystem contains the commands to control the LF sweep of SOURce2. LF-Sweeps are
activated by command SOUR2:MODE SWE. Sweeps are triggered on principle.
Command Parameters Default
Unit
Remark
:SOURce2
:SWEep
[:FREQuency]
:DWELl
:MODE
:RUNNinng?
10 ms...5 s
AUTO | MANual | STEP
s
Query only
:SPACing
:STEP
[:LINear]
:LOGarithmic
LINear | LOGarithmic
0...1 MHz
0.01 PCT...100 PCT
Hz
PCT
:SOURce2:SWEep[:FREQuency]
The commands to set the frequency sweeps are under this node. Keyword [:FREQuency] can be
omitted. Then the commands are SCPI-compatible unless stated otherwise (see examples).
:SOURce2:SWEep[:FREQuency]:DWELl 10 ms to 5 s
The command sets the time per frequency step (dwell).
Example: :SOUR2:SWE:DWEL 20ms *RST value is 15 ms
:SOURce2:SWEep[:FREQuency]:MODE AUTO | MANual | STEP
The command specifies the run of the sweep.
AUTO Each trigger triggers exactly one entire sweep cycle.
STEP Each trigger triggers only one sweep step (single-step mode). The frequency
increases by the value defined under :SOUR2:SWE:STEP.
Example: :SOUR2:SWE:MODE AUTO *RST value is AUTO
:SOURce2:SWEep[:FREQuency]:SPACing LINear | LOGarithmic
The command selects whether the steps have linear or logarithmic spacings.
Example: :SOUR2:SWE:SPAC LOG *RST value is LIN