R&S®SML / R&S®SMV03 SOURce:PULSe
1090.3123.12 E-6 6.31
SOURce:PULSe Subsystem
This subsystem contains the commands to set the pulse generator (Option R&S SML-B3). The pulse
generation is triggered on principle, with the trigger certainly being able to be set to "free run" using
TRIG:PULS:SOUR AUTO as well.
Command Parameters
Default
Unit
Remark
[:SOURce]
:PULSe
:DELay
:DOUBle
:DELay
:STATe
:PERiod
:WIDTh
20 ns to 1.3 s
60 ns to 1.3 s
ON | OFF
100 ns to 85 s
20 ns to 1.3 s
s
s
s
s
Option R&S SML-B3
[:SOURce]:PULSe:DELay 20 ns to 1.3 s
The command specifies the time from the start of the period to the first edge of the pulse. Due to
the construction of the instrument, this parameter is set to 0 if [:SOUR]:PULS:DOUB:STAT is
set to ON. The old value is activated again as soon as the double pulse has been switched off.
Example: :SOUR:PULS:DEL 10us *RST value is 1 Ts
[:SOURce]:PULSe:DOUBle
The commands to check the second pulse are under this node. If [:SOUR]:PULS:DOUB:STAT
is set to ON, a second pulse whose width is identical to the first pulse is generated in every period.
[:SOURce]:PULSe:DOUBle:DELay 60 ns to 1.3 s
The command sets the delay time from the start of the pulse period to the first edge of the second
pulse.
Example: :SOUR:PULS:DOUB:DEL 10us *RST value is 1 Ts
[:SOURce]:PULSe:DOUBle:STATe ON | OFF
The command switches the second pulse on or off.
ON The second pulse is switched on.
Parameter [:SOUR]:PULS:DEL is set to 0 and cannot be changed. WIDTh >
(PULS:PER - PULS:DOUB:DEL)/2 results in error message -221, "Settings conflict".
OFF The second pulse is switched off.
Example: :SOUR:PULS:DOUB:STAT OFF *RST value is OFF
[:SOURce]:PULSe:PERiod 100 ns to 85 s
The command sets the pulse period.
The pulse period is the reciprocal value of the pulse frequency, thus this command is coupled to
command [:SOUR]:PULM:INT:FREQ.