R&S®SML / R&S®SMV03 General
1090.3123.12 10.1 E-6
10 Performance Test
General
• The rated specifications of the signal generator are checked after a warm-up time of at least 15
minutes. Recalibration of the instrument is not necessary, with the exception of the FM offset
calibration.
• A defined default state is set prior to each measurement by pressing the PRESET key.
• The values in the following sections are not guaranteed. Only the specifications in the data sheet are
binding.
• The values in the data sheet are guaranteed limits. Due to measurement errors, these limits must be
extended by the tolerances of the measuring instruments used in the performance test.
• The maximum frequency f
max
that can be set for each of the different R&S SML models is shown in
Table 10-1. This should be borne in mind when choosing test equipment.
Table 10-1 Maximum frequencies
Model R&S
SML01
R&S
SML02
R&S
SML03
R&S
SMV03
f
max
[MHz] 1100 2200 3300 3300