R&S®SML / R&S®SMV03 General
1090.3123.12 10.1 E-6
10 Performance Test
General
The rated specifications of the signal generator are checked after a warm-up time of at least 15
minutes. Recalibration of the instrument is not necessary, with the exception of the FM offset
calibration.
A defined default state is set prior to each measurement by pressing the PRESET key.
The values in the following sections are not guaranteed. Only the specifications in the data sheet are
binding.
The values in the data sheet are guaranteed limits. Due to measurement errors, these limits must be
extended by the tolerances of the measuring instruments used in the performance test.
The maximum frequency f
max
that can be set for each of the different R&S SML models is shown in
Table 10-1. This should be borne in mind when choosing test equipment.
Table 10-1 Maximum frequencies
Model R&S
SML01
R&S
SML02
R&S
SML03
R&S
SMV03
f
max
[MHz] 1100 2200 3300 3300