Storing/Calling of Instrument Settings R&S®SML / R&S®SMV03
1090.3123.12 E-6 3.14
Storing/Calling of Instrument Settings
(SAVE / RECALL)
50 complete instrument settings can be stored in memory locations 1 to 50.
Operating Steps Explanations
SAVE
DATA INPUT
x1
ENTER
1 2
dB(m)
Store current instrument setting in
memory location 12.
RCL
DATA INPUT
x1
ENTER
12
dB(m)
Call instrument setting of memory
location 12.
The digital display during a save or recall entry is faded in a window.
If an instrument setting is stored in which a sweep was switched on, the sweep is started using the
recall.
The parameter Exclude From Recall in the Frequency and Level-Level menus determines whether the
saved RF frequency and RF level are loaded when an instrument setting is loaded, or whether the
current settings are maintained.
Store IEC-bus command: "*SAV 12"
Call IEC-bus command: "*RCL 12"
Note: The contents of lists, as they are used for user correction (Ucor), is not saved in the SAVE
memory. It is stored under the respective list name and can be called. If instrument settings
are called which go back to list data such as level setting using Ucor, the current list
contents is used. If this has been altered, it is not identical to the list contents at the point of
storing any more.