SOURce:PULM R&S®SML / R&S®SMV03
1090.3123.12 E-6 6.30
SOURce:PULM Subsystem
This subsystem contains the commands to control the pulse modulation (Option R&S SML-B3) and to
set the parameters of the modulation signal. The internal pulse generator is set in the :SOURce:PULSe
subsystem.
Command Parameters Default
Unit
Remark
[:SOURce]
:PULM
:EXTernal
:POLarity
:SOURce
:STATe
NORMal | INVerse
INTernal | EXTernal
ON | OFF
Option R&S SML-B3
[:SOURce]:PULM:EXTernal
The commands to control the input socket for the external pulse generator are under this node.
[:SOURce]:PULM:POLarity NORMal | INVerse
The command specifies the polarity between modulating and modulated signal.
NORMal The RF signal is suppressed during the interpulse period.
INVerse The RF signal is suppressed during the pulse.
Example: :SOUR:PULM:POL INV *RST value is NORM
[:SOURce]:PULM:SOURce EXTernal | INTernal
The command selects the source of the modulating signal.
INTernal Internal pulse generator.
EXTernal Signal fed externally.
Example: :SOUR:PULM:SOUR INT *RST value is INT
[:SOURce]:PULM:STATe ON | OFF
The command switches on or off the pulse modulation.
Example: :SOUR:PULM:STAT ON *RST value is OFF