R&S®SML / R&S®SMV03 SOURce:POWer
1090.3123.12 E-6 6.29
[:SOURce]:POWer:STARt -1
40 dBm to P
max
(+29 dBm with R&S SML-B10)
The command sets the staring value for a level sweep. STARt may be larger than STOP, then the
sweep runs from the high to the low level (As to specified range, cf. :POW).
Example: :SOUR:POW:STAR -20 *RST value is -30 dBm
[:SOURce]:POWer:STOP -140 dBm to P
max
(+29 dBm with R&S SML-B10)
The command sets the final value for a level sweep. STOP may be smaller than STARt. (As to
specified range, cf. :POW).
Example: :SOUR:POW:STOP 3 *RST value is -10dBm
[:SOURce]:POWer:STEP[:INCRement] 0.1 to 10 dB
The command sets the step width with the level setting if UP and DOWN are used as level
values. The command is coupled to Knob Step in the manual control, i.e., it also specifies the
step width of the shaft encoder.
Only dB is permissible as a unit here, the linear units (V, W etc.) are not permitted.
Example: :SOUR:POW:STEP:INCR 2 *RST value is 1dB