R&S®SML / R&S®SMV03  Test sequence 
1147.7621.12 10.43  E-6 
Level  
Level frequency response and linearity  
Test method for level in measurement range of power meter (up to approx. - 22 dBm) 
 
Test equipment   See Performance test R&S SML03 / R&S SMV03. 
Test setup  
See Performance test R&S SML03 / R&S SMV03. 
 
Settings on  
R&S SMV03 
- VECTORMOD STATE IQCW 
- FREQ Test frequency level accuracy   
- Test frequencies: 5 MHz, 5,1 MHz, 76 MHz, 77 MHz, 151 MHz, 
255 MHz, 302 MHz, 605 MHz, 606 MHz, 725 MHz, 970 MHz, f
max
,
1210.5 MHz, 1211 MHz, 1818 MHz, 2000 MHz, 2200 MHz, 2800 
MHz, 3000 MHz, 3300 MHz 
- LEVEL Test level  1 level accuracy  
- Test level 1:  8 dBm, 3.1 dBm, 3 dBm, -2 dBm, -7 dBm,  
-12 dBm, -17 dBm, -21.9 dBm, -22.0 dBm 
 
Settings on powermeter   See Performance test R&S SML03 / R&S SMV03 
Measurement   See Performance test R&S SML03 / R&S SMV03. 
 
Measurement procedure for low levels (>-115 dBm) 
Caution:  The precondition for correct measurement is that the used components are wholly 
RF-shielded. 
 
Test Method see Performance test R&S SML03 / R&S SMV03. 
 
Settings on  R&S SMV03  -  VECTORMOD STATE IQCW 
- Test frequencies  >=5MHz  
 
Setting time  
Test setup   Test setup 2 (see Performance test R&S SML03 / R&S SMV03) 
 
Test method   See Performance test R&S SML03 / R&S SMV03 
 
Preparing measurement  See Performance test R&S SML03 / R&S SMV03 
 
Settings on R&S SMV03  -  VECTORMOD STATE IQCW 
- FREQ Test frequency setting time  
MeasurementMessung  See Performance test R&S SML03 / R&S SMV03 
 
Non-interrupting level setting (ATTENUATOR FIXED)  
Tset setup    Test setup 1 with spectrum analyzer (Performance test R&S 
SML03 / R&S SMV03) 
Settings on R&S SMV03  -  VECTORMOD STATE IQCW 
- FREQ Test frequencies  ATT-FIX  
- Test frequencies: 100 kHz, 5,1 MHz, f
max
, 1211 MHz.  
2200 MHz, 3300 MHz 
- LEVEL  3.1 dBm  
- LEVEL LEVEL  ATTENUATOR MODE  FIXED 
 
Measurement   See Performance test R&S SML03 / R&S SMV03