Test sequence R&S®SML / R&S®SMV03
1147.7621.12 10.42 E-6
Nonharmonic suppression
Test setup Test setup 1 with spectrum analyzer (see Performance test R&S
SML03 / R&S SMV03)
Settings on R&S SMV03
- UTILITIES REF OSC SOURCE EXTERNAL
- LEVEL 10 dBm
- VECTORMOD STATE IQCW
- FREQ Test frequecy of nonharmonics Test frequencies of
nonharmonics: 250.0MHz, 250.0000001MHz
Setting on spectrum analyzer - Reference level = test level + 3 dB, 10 dB/div
- Test frequency = 200 MHz, 800 MHz, 1600 MHz, 2400 MHz, 3200
MHz
- Span = 1 MHz
- Switch on average: 5 samples
Measurement See Performance test R&S SML03 / R&S SMV03.
Evaluation
See Performance test R&S SML03 / R&S SMV03.
Broadband noise
Test setup
Test setup 3 (see Performance test R&S SML03 / R&S SMV03)
Settings on
R&S SMV03
- UTILITIES REF OSC SOURCE EXTERNAL
- VECTORMOD STATE IQCW
- LEVEL 0 dBm (or . level according to mixer specification)
- FREQUENCY 1 GHz (or any test frequency)
Test method See Performance test R&S SML03 / R&S SMV03.
Measurement See Performance test R&S SML03 / R&S SMV03
Evaluation See Performance test R&S SML03 / R&S SMV03.