R&S®SML / R&S®SMV03 STATus
1090.3123.12 E-6 6.45
STATus System
This system contains the commands for the status reporting system (c.f. Section "Status Reporting
System"). STATus:OPERation register and STATus:QUEStionable register are not implemented. *RST
has no influence on the status registers.
Command Parameters Default
Unit
Remark
:STATus
:PRESet
:QUEue
[:NEXT]?
No query
Query only
:STATus:PRESet
The command resets the edge detectors and ENABle parts of all registers to a defined value. All
PTRansition parts are set to FFFFh, i.e., all transitions from 0 to 1 are detected. All NTRansition
parts are set to 0, i.e., a transition from 1 to 0 in a CONDition bit is not detected.
Example: :STAT:PRES
:STATus:QUEue [:NEXT]?
The command queries the entry that has been in the error queue for the longest time and thus
deletes it. Positive error numbers denote errors specific of the instrument, negative error numbers
error messages specified by SCPI (see Chapter 5). If the error queue is empty, 0, "No error", is
returned. The command is identical to SYST:ERR?
Example: STAT:QUE:NEXT? Answer: 221, "Settings conflict"