R&S®SML / R&S®SMV03 RF Level
1090.3123.12 4.7 E-6
User Correction (Ucor)
The "User correction" function can be used to create and activate lists in which level correction values
are assigned to arbitrary RF frequencies.
Up to 10 lists with a total of 160 correction values can be compiled. For frequencies not included in the
list, level correction values are determined by interpolation based on the nearest correction values.
When user correction is switched on, Ucor (user correction) is displayed in the header field in addition to
the level. The RF output level is the sum of both values.
Level + Ucor = output level
If an offset is selected at the same time, the displayed level value is the difference between the
amplitude and the offset entered in the Level menu.
Amplitude offset = level
User correction is active in all operating modes when switched on.
Menu selection: Level - UCor
Fig. 4-7 Level - UCor menu
State
Switching on/off user correction
IEC/IEEE-bus command :SOUR:CORR ON
Select List
Selection of a list or generation of a new list
(see Chapter 3, Section "List Editor")
IEC/IEEE-bus command :SOUR:CORR:CSET "UCOR1"
Delete List
Deletion of a list (see Chapter 3, Section "List Editor")
IEC/IEEE-bus command :SOUR:CORR:CSET:DEL "UCOR2"
Edit List
Selection of editing mode for modifying a selected list
(see Chapter 3, Section "List Editor")
IEC/IEEE-bus commands
:SOUR:CORR:CSET:DATA:FREQ 105MHz, 107MHz,...
:SOUR:CORR:CSET:DATA:POW 1dB, 0.9dB, 0.8dB,...