R&S®SML / R&S®SMV03  RF Level 
1090.3123.12 4.7  E-6 
User Correction (Ucor)  
The "User correction" function can be used to create and activate lists in which level correction values 
are assigned to arbitrary RF frequencies. 
 
Up to 10 lists with a total of 160 correction values can be compiled. For frequencies not included in the 
list, level correction values are determined by interpolation based on the nearest correction values. 
 
When user correction is switched on, Ucor (user correction) is displayed in the header field in addition to 
the level. The RF output level is the sum of both values. 
Level + Ucor = output level 
If an offset is selected at the same time, the displayed level value is the difference between the 
amplitude and the offset entered in the Level menu. 
Amplitude  offset = level 
 
User correction is active in all operating modes when switched on. 
 
Menu selection:   Level - UCor 
Fig. 4-7  Level - UCor menu  
State 
Switching on/off user correction 
IEC/IEEE-bus command  :SOUR:CORR ON 
Select List 
Selection of a list or generation of a new list   
(see Chapter 3, Section "List Editor") 
IEC/IEEE-bus command  :SOUR:CORR:CSET "UCOR1" 
Delete List 
Deletion of a list (see Chapter 3, Section "List Editor") 
IEC/IEEE-bus command  :SOUR:CORR:CSET:DEL "UCOR2" 
Edit List 
Selection of editing mode for modifying a selected list   
(see Chapter 3, Section "List Editor") 
IEC/IEEE-bus commands 
  :SOUR:CORR:CSET:DATA:FREQ 105MHz, 107MHz,...
:SOUR:CORR:CSET:DATA:POW 1dB, 0.9dB, 0.8dB,...